The CP 210-R measures the differential pressure of air (and neutral gases), and is ideal for use in industrial and pharmaceutical applications.

Measuring range: CP 211: -100 to +100 Pa, CP 212: -1,000 to +1,000 Pa, and CP 213: -10,000 to +10,000 Pa
Solenoid valve for auto-calibration (CP211/212)
0-5/10 V or 0/4-20 mA outputs
Two relays
Automatic autozero
Quick and easy mounting with the 1/4 turn system with wall-mount plate

    Measure details

    • Units
      °F, °C, Pa, kPa, mbar, hPa, inWg, mmH₂O, mmHg, daPa
    shadow

    Technical details

    • Parameters
      Differential pressure
      Temperature
    • Displays
      No
      2-line display
    • IP Protection
      IP 65
    • Standards
      UKCA
      2011/65/EU RoHS II
      2012/19/EU WEEE
      2014/30/EU EMC
      2014/35/EU Low Voltage
    • Power supply
      24 Vac/Vdc ou 100-240 Vac
    • Outputs
      2 x 4-20 mA or 2 x 0-20 mA or 2 x 0-5 V or 2 x 0-10 V (4 fils)
    • Relay
      2 RCR relay 3A/230 Vac
    • Ethernet communication
      No
    • ModBus network
      No
    shadow

    Instrument Dimensions

    • Length
      125 mm
    • Width
      59.2 mm
    • Height
      115 mm
    • Weight
      340 g
    shadow

    Operating temperature

    • Min
      -10.00 °C
    • Max
      50.00 °C
    shadow

    Storage temperature

    • Min
      -10.00 °C
    • Max
      70.00 °C
    shadow